Specifications

Method of Detection

Long-life capacitance sensor

Ranges Available

-112°F to +68°F (-80°C to +20°C) Dew Point

Resolution

1°F (1°C) Dew Point

Accuracy & Repeatability

±1°C from -60 to +20°C Dew Point; ±2°C from -80 to -60°C
±1°F from -76 to +68°F Dew Point; ±2°F from -112 to -76°F

Drift

±2% of full scale per month maximum

Response Time (T-90)

Less than 30 seconds at higher dew points; lower dew points may take up to 1 minute (sensor must be in pre-dried state)

Ambient Temperature Range

-18 to 50 ºC
0 to 120 ºF

Linearity

±2°F/C over measurement range

Size & Weight

BT Style - Approx. 9" H x 8" W x 10" D @ 8 lbs (23 x 20 x 25 cm @ 3.6 kg)
K Style - Approx. 14" L x 6" H x 10 1/2" D @ 12 lbs (35 x 15 x 26 cm @ 5.5 kg)

Power

115 VAC, 60 Hz for recharging
220 VAC 50 Hz available

Output Options

0-1 V or 4-20 mA

Alarms

High or low dew point alarms with LED (optional)

Hazardous Area Rating Options

General Purpose, Non-Hazardous


Applications
    For checking lower range dew points of air, hydrogen/nitrogen atmosphere furnace gases or other dry, oil-free process gases.



Features
  • Low range of -112 to +68°F (or -80 to 20°C) Dew Point
  • Easy to operate
  • Fast warm-up and response
  • Use on any clean, non-corrosive industrial gases or air
  • Long-life dew point sensor
  • May be calibrated in the field
  • Large LCD digital readout with backlight
  • Configurations available for wide range of sample pressures
  • No external adjustments to make



Options
  • Recorder outputs of 0-1 V or 4-20 mA
  • Rugged "K" style weathertight suitcase enclosure
  • high or low dew point alarm




Calibration
  • On analyzed dry nitrogen (N2) for low setting
  • On known, current ambient air dew point for span
  • Alternatively, the analyzer may be returned to Nova for a certified calibration




Unique Applications
    The 206 Series should not be used to measure dew points when trace amounts of sulfur trioxide (SO3), chlorine (Cl2), hydrogen chloride (HCl), or ammonia (NH3) are present.
Download
 
2010©Copyright  Combine instruments Co.,ltd allrights revered
USA:17888 67TH COURT NORTH LOXAHATCHEE FL 33470 US Post:FL33470
CHINA: No.40,JinFengXi Road,National Hi-Tech Industry Development Zone,ZhuHai,GuangDong,China
Tel:+0086 756 3895799 Fax:0086 756 3892788 E-mail:sales@combine-inst.com 

COMBINE INSTRUMENTS   Batching System  Loss-in-weight-feeder 
粤ICP备16107310号