Method of Detection |
 |
Customer replaceable electrochemical NO sensor
|
 |
Ranges Available |
 |
0-2000 PPM NOx (as NO)
|
 |
Resolution |
 |
1 PPM NO
|
 |
Accuracy & Repeatability |
 |
Better than 2% full scale
|
 |
Drift |
 |
Within 1% full scale per 8 hours of continuous operation
|
 |
Response Time (T-90) |
 |
20-30 seconds for 90% of step change
|
 |
Ambient Temperature Range |
 |
12 to 49 ºC 55 to 120 ºF
|
 |
Linearity |
 |
±2% of full scale
|
 |
Size & Weight |
 |
14" L x 10 1/2" W x 6" H @ 12 lbs (36 x 27 x 15 cm @ 5.5 kg)
|
 |
Power |
 |
AC/DC operation, 115 VAC 60 Hz for recharging (220 VAC 50 Hz available)
|
 |
Output Options |
 |
4-20 mA or 0-1 VDC
|
 |
Alarms |
 |
NOx alarms with LED
|
 |
Hazardous Area Rating Options |
 |
General Purpose, Non-Hazardous
|
 |
| |
|
Analysis of oxides of nitrogen (NOx) as nitric oxide (NO) for boiler, furnace, or engine exhaust.
| |
|
- Rugged design that is easy to operate and maintain
- Fast warm-up and response
- Disposable, long life electrochemical NO sensor
- Digital readout meter with backlight
- Rechargeable battery operation
- Built in sample pump, filter and flowmeter
- Active condensate removal
- Rapid recovery on NOx overdose
- Stainless steel probe with sample hose
| |
|
- Recorder outputs of 0-1 V or 4-20 mA
- Stack temperature readout (312T)
- Special sampling probes
- Sample pre-cooler
- Suitcase (K) style cabinet available
- AC power only operation
- NOx alarms with LED
- Detachable/portable datalogger
| |
|
- Air for zero
- Analyzed calibration gas mixture of PPM NO in air for span.
| | |
|
|
|