Method of Detection |
|
Temperature compensated thermal conductivity (T/C) cell
|
|
Ranges Available |
|
Range 1 : 0-100% H2 in Air Range 2 : 0-100% H2 in CO2 Range 3 : 0-100% Air in CO2
|
|
Resolution |
|
0.1% of gas measured
|
|
Accuracy & Repeatability |
|
±1% F.S except in 85-100% H2 in Air range, which is within 0.5% absolute H2
|
|
Drift |
|
0-100% H2 in Air range is ± 0.4% per week H2 in CO2 or Air in CO2 2% F.S. per week maximum drift
|
|
Response Time (T-90) |
|
10-15 seconds to 90% step change - not including sample transport time
|
|
Ambient Temperature Range |
|
0 to 50 ºC 32 to 120 ºF
|
|
Linearity |
|
±1.0% of F.S. on H2 in Air range. ±2% of F.S. in H2 or Air in CO2 ranges
|
|
Size & Weight |
|
9 1/2" L x 7" W x 6 1/2" H @ 12 lbs (24 x 17 x 18 cm @ 3.6 kg)
|
|
Power |
|
AC/DC Operation. 115 VAC, 60 Hz for recharging (220 VAC, 50 Hz available)
|
|
Output Options |
|
4-20 mA or 0-1 VDC
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
For checking purity of hydrogen (H2) in H2-cooled generators and synchronous converters. The Model 380 Series will also monitor the safe purging of H2 during shut down.
| |
|
- Rugged design that is easy to operate
- Fast warm up and response
- Monitors 0-100% H2 in air, 0-100% H2 in CO2, 0-100% Air in CO2
- Rugged long life thermal conductivity cell
- Digital readout meter with backlight
- Rechargeable battery operation
- Modular layout that is easy to maintain
- Built in flow meter, flow control valve and pump
- Built in pressure regulator prevents over pressuring of pump and sensor
- Weatherproof (WP) cabinet with clear lexan cover
- Sample pressure ranges 0.5 PSI to 125 PSI
| |
|
- Recorder outputs of 0-1 V or 4-20 mA
- Suitcase (K) style cabinet available
| |
|
- Range 1 : Air for Zero, 100% H2 for Span
- Range 2 : 100% CO2 for Zero, 100% H2 for Span
- Range 3 : 100% CO2 for Zero, Air for Span.
| | |
|
|
|