Method of Detection |
 |
Temperature compensated thermal conductivity (T/C) cell
|
 |
Ranges Available |
 |
0-100% H2 in a binary gas mixture
|
 |
Resolution |
 |
0.1% of H2
|
 |
Accuracy & Repeatability |
 |
±2% of full scale
|
 |
Drift |
 |
±1% of full scale max. per day (after calibration)
|
 |
Response Time (T-90) |
 |
10-15 seconds to 90% step change - not including sample transport time
|
 |
Ambient Temperature Range |
 |
12 to 50 ºC 55 to 120 ºF
|
 |
Linearity |
 |
±2% of F.S.
|
 |
Size & Weight |
 |
9 1/2" L x 7" W x 6 1/2" H @ 8 lbs (24 x 17 x 18 cm @ 3.6 kg)
|
 |
Power |
 |
AC/DC Operation. 115 VAC, 60 Hz for recharging (220 VAC, 50 Hz available)
|
 |
Output Options |
 |
0-1 VDC 4-20 mA
|
 |
Alarms |
 |
H2 alarm with LED (optional)
|
 |
Hazardous Area Rating Options |
 |
General Purpose, Non-Hazardous
|
 |
| |
|
For analysis of hydrogen (H2) in a binary gas mixture in process gases such as H2 in air, H2 in nitrogen (N2), H2 in carbon dioxide (CO2), H2 in argon (Ar), H2 in oxygen (O2), etc.
| |
|
- Rugged design that is easy to operate
- Fast warm up and response
- Long life thermal conductivity cell that provides accurate and stable readings
- Digital meter readout with backlight
- Modular layout that is easy to maintain
- Rechargeable 'gel cell' battery operated
- Built in sample pump or flow regulator
- Weatherproof (WP) cabinet with clear lexan cover
| |
|
- Recorder outputs of 0-1 V or 4-20 mA
- Sample pre-cooler for hot samples
- Condensate removal for wet applications
- Suitcase (K) style cabinet available
- AC power only operation
- H2 alarm with LED
- Detachable/portable data logger
| |
|
- Ambient air for zero
- Gas cylinder of known H2 for span
| | |
|
|
|