Method of Detection |
|
Microprocessor based infrared detector for CO2
|
|
Ranges Available |
|
0-3000, 0-5000 or 0-10,000 PPM; 0-2%, 0-5%, 0-10%, 0-20%, 0-50% or 0-100% CO2
|
|
Resolution |
|
0.1% CO2 on % versions; 10 PPM CO2 on PPM versions
|
|
Accuracy & Repeatability |
|
Within ±2% full scale
|
|
Drift |
|
Within 1% full scale per 8 hours of continuous operation
|
|
Response Time (T-90) |
|
30 - 40 seconds
|
|
Ambient Temperature Range |
|
4 to 49 ºC 40 to 120 ºF
|
|
Linearity |
|
Better than 1.0% of full scale
|
|
Size & Weight |
|
11.5" L x 8" W x 7 1/4" D @ 8 lbs (29 x 20 x 18 cm @ 3.6 kg)
|
|
Power |
|
AC/DC Operation, 115 VAC 60 Hz for recharging
|
|
Output Options |
|
4-20 mA or 0-1 VDC
|
|
Alarms |
|
High or low CO2 alarms with LED (optional)
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
For checking carbon dioxide (CO2) content in mushroom growing rooms, incubators, natural gas, controlled atmosphere storage rooms, welding gas mixtures, and other process applications.
| |
|
- Rugged design that is easy to operate
- Fast warm-up and response
- Solid state infrared detector
- Digital readout meter with backlight
- Modular layout that is easy to maintain
- Rechargeable battery operation
- Built in sample pump, filter, and flowmeter
- Weatherproof (WP) cabinet with clear Lexan cover
| |
|
- Recorder outputs of 0-1 V or 4-20 mA
- Sample pre-cooler for hot samples
- Condensate removal for wet applications
- Suitcase (K) style cabinet available
- AC power only operation
- CO2 alarms with LED
- Data logger
| |
|
- Air for zero (except low range models which have a CO2 scrubber for zeroing).
- Analyzed CO2 cal gas for span.
| | |
|
|
|