|
|
Method of Detection |
|
Customer replaceable electrochemical 'fuel cell' CO sensor. Expected life 1 to 3 years depending on typical measurement concentrations.
|
|
Ranges Available |
|
Any range from 0-100 PPM to 0-10000 PPM C0
|
|
Resolution |
|
1 PPM CO
|
|
Accuracy & Repeatability |
|
±1% of full scale
|
|
Drift |
|
Less than 1% of full scale per month
|
|
Response Time (T-90) |
|
6-8 seconds to 90% step change
|
|
Ambient Temperature Range |
|
32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package
|
|
Linearity |
|
±1% of full scale
|
|
Size & Weight |
|
Dimensions will vary depending on enclosure style and options required
|
|
Power |
|
115 VAC, 60 Hz 220 VAC, 50 Hz
|
|
Output Options |
|
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
|
|
Alarms |
|
High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
|
|
Analyzer Type |
|
CO Analyzer
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
For continuous analysis of up to 0-10,000 PPM carbon monoxide (CO) in process gas streams using an electrochemical detector.
| |
|
- Best value continuous low-range CO analyzer
- Long life (1-3 years) electrochemical CO sensor
- Bright digital readout
- Fast response (T90 - 6 to 8 seconds)
- Microprocessor control for push button calibration
- Easy-to-maintain modular layout
- Built in sample pump or pressure regulator
- Sensor mounted in temperature controlled heated box to improve accuracy
- 4-20 mA outputs
| |
|
- Hi/Low gas, low flow, and diagnostic alarms available
- Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
- Sample conditioning systems available for removal of moisture, acidic gas, and dust
- Cabinet purge system available for use in hazardous areas
- One-step 'cal-now' button with touch screen LCD display
- Full automatic calibration with touch screen LCD display
- Cold weather package for operation to -5°F (-20°C)
- Cabinet coolers can be fitted to most models
- High temperature filters and probes
| |
|
- Nitrogen for zero.
- Analyzed CO calibration gas for span.
| | |
|
|
|