Method of Detection |
 |
Low range electrochemical O2 sensor
|
 |
Ranges Available |
 |
0-100 PPM to 0-9999 PPM O2
|
 |
Resolution |
 |
1 PPM
|
 |
Accuracy & Repeatability |
 |
±2% of reading
|
 |
Drift |
 |
±1% of full scale per week
|
 |
Response Time (T-90) |
 |
O2 sensor may require an initial purge down from air to low PPM O2, which may last up to 6 hours. After purge down, T-90 response from 1000 to 100 PPM is 30 seconds.
|
 |
Ambient Temperature Range |
 |
0-50 ºC 32-122 ºF
|
 |
Linearity |
 |
±2% of full scale
|
 |
Size & Weight |
 |
14" W x 6" H x 6 1/2" D @ 12 lbs. (35 x 15 x 17 cm @ 5.5 kg)
|
 |
Power |
 |
AC/DC operation. 115 VAC 60 Hz for recharging (Other voltages available)
|
 |
Output Options |
 |
0-1V or 4-20mA (optional)
|
 |
Alarms |
 |
One high or low O2 alarm (optional)
|
 |
Hazardous Area Rating Options |
 |
General Purpose, Non-Hazardous
|
 |