Method of Detection |
 |
Customer replaceable electrochemical 'fuel cell' oxygen sensor, expected life 5 years
|
 |
Ranges Available |
 |
0-30% O2
|
 |
Resolution |
 |
0.1% O2
|
 |
Accuracy & Repeatability |
 |
±2% of full scale
|
 |
Drift |
 |
Less than 1% of full scale per month
|
 |
Response Time (T-90) |
 |
6-8 seconds to 90% step change
|
 |
Ambient Temperature Range |
 |
32 to 122°F (0 to 50°C), may be operated at lower temperature with auxiliary heater
|
 |
Linearity |
 |
±2% of full scale
|
 |
Size & Weight |
 |
10" H x 8" W x 6" D (25.4 x 20.3 x 15.2 cm) @ 6 lbs. (2.7 kg)
|
 |
Power |
 |
115 VAC, 60 Hz 12 VDC, 60 Hz 220 VAC, 50 Hz operation 24 VDC, 60 Hz
|
 |
Output Options |
 |
0-1 VDC 4-20 mA
|
 |
Alarms |
 |
High and low SPDT contacts with 10 A, 115 VAC rating; Two alarm levels standard; 'Sonalert' audible alarm with reset button; Indicator light for each alarm level; Strobe light and 'fail safe' alarms available
|
 |
Analyzer Type |
 |
O2 Gas Analyzer
|
 |
| |
|
To monitor confined spaces, labs, control rooms, tunnels, oxygen plants, etc. for oxygen (O2) deficiency or excess.
| |
|
- Best value high accuracy O2 monitor
- Bright digital readout
- 5 year life electrochemical O2 sensor
- Two levels of O2 alarm with relay contacts and 'Sonalert' audible alarm with 'Acknowledge' button
- Easy to use and calibrate
- Fast response (T90 - 6 to 8 seconds)
- Corrosion resistant NEMA 12 enclosure
- 4-20 mA or 0-1 VDC recorder output
- Modular layout that is easy to maintain
| |
|
- 'Fail Safe' alarms available
- Low flow alarm on pumped sample models
- NEMA 4X enclosure
- Strobe light
| |
|
- Although calibration of zero should not be necessary, it can be checked by flowing 1 LPM of nitrogen through the analyzer
- Analyzed gas or ambient air with 20.9% O2 for span
| | |
|
|
|