Language:       
High-quality products
&
professional services
 
  About us
  Products
  News center
  Support & serivces
  Solution & Application
  Service Network
  Contact us
  link



Specifications

Method of Detection

Customer replaceable electrochemical 'fuel cell' oxygen sensor, expected life 5 years

Ranges Available

0-30% O2

Resolution

0.1% O2

Accuracy & Repeatability

±2% of full scale

Drift

Less than 1% of full scale per month

Response Time (T-90)

6-8 seconds to 90% step change

Ambient Temperature Range

32 to 122°F (0 to 50°C), may be operated at lower temperature with auxiliary heater

Linearity

±2% of full scale

Size & Weight

10" H x 8" W x 6" D (25.4 x 20.3 x 15.2 cm) @ 6 lbs. (2.7 kg)

Power

115 VAC, 60 Hz
12 VDC, 60 Hz
220 VAC, 50 Hz operation
24 VDC, 60 Hz

Output Options

0-1 VDC
4-20 mA

Alarms

High and low SPDT contacts with 10 A, 115 VAC rating; Two alarm levels standard; 'Sonalert' audible alarm with reset button; Indicator light for each alarm level; Strobe light and 'fail safe' alarms available

Analyzer Type

O2 Gas Analyzer


Applications
    To monitor confined spaces, labs, control rooms, tunnels, oxygen plants, etc. for oxygen (O2) deficiency or excess.




Features
  • Best value high accuracy O2 monitor
  • Bright digital readout
  • 5 year life electrochemical O2 sensor
  • Two levels of O2 alarm with relay contacts and 'Sonalert' audible alarm with 'Acknowledge' button
  • Easy to use and calibrate
  • Fast response (T90 - 6 to 8 seconds)
  • Corrosion resistant NEMA 12 enclosure
  • 4-20 mA or 0-1 VDC recorder output
  • Modular layout that is easy to maintain




Options
  • 'Fail Safe' alarms available
  • Low flow alarm on pumped sample models
  • NEMA 4X enclosure
  • Strobe light




Calibration
  • Although calibration of zero should not be necessary, it can be checked by flowing 1 LPM of nitrogen through the analyzer
  • Analyzed gas or ambient air with 20.9% O2 for span
Download
 
2010©Copyright  Combine instruments Co.,ltd allrights revered
USA:17888 67TH COURT NORTH LOXAHATCHEE FL 33470 US Post:FL33470
CHINA: No.40,JinFengXi Road,National Hi-Tech Industry Development Zone,ZhuHai,GuangDong,China
Tel:+0086 756 3895799 Fax:0086 756 3892788 E-mail:sales@combine-inst.com 

COMBINE INSTRUMENTS   Batching System  Loss-in-weight-feeder 
粤ICP备16107310号