产品名称:
340WP便携式分析仪
|

|
Method of Detection |
 |
Long-life electrochemical O2 sensor, temperature-compensated thermal conductivity (T/C) H2 cell cannot be burned out due to loss of flow or changing gases
|
 |
Ranges Available |
 |
0-100 to 0-9999 PPM O2, 0-40.0% H2 (Model 340L) 0-2.00% and 0-25.0% O2, 0-40.0% H2 (Model 341) 0-25.0% O2, 0-40.0% H2 (Model 340)
|
 |
Resolution |
 |
0.1% on % ranges; 1 PPM on PPM O2 range
|
 |
Accuracy & Repeatability |
 |
±2% of full scale
|
 |
Drift |
 |
±1% of full scale per day
|
 |
Response Time |
 |
Less than 10 seconds to 90% step change
|
 |
Ambient Temperature Range |
 |
12-50 ºC 55-120 ºF
|
 |
Linearity |
 |
±2% of full scale
|
 |
Size & Weight |
 |
11 1/2" L x 8" W x 7 1/4" H @ 8 lbs ( 29 x 20 x 18 cm @ 3.6kg)
|
 |
Power |
 |
AC/DC operation, 115 VAC 60 Hz for recharging (220 VAC 50Hz available)
|
 |
Output Options |
 |
4-20 mA or 0-1VDC
|
 |
Alarms |
 |
O2 and H2 alarms with LED (optional)
|
 |
Hazardous Area Rating Options |
 |
General Purpose/Non-hazardous
|
 |
| |
|
Analysis of oxygen/hydrogen (O2/H2) in exothermic furnace atmospheres for copper, brass or steel anealing, neutral heating, sintering, glass metal beds or oxide coating of steel.
| |
|
Rugged modular design that is easy to operate and maintain
Fast warm-up and response
Long life electrochemical O2 sensor easily detects air infiltration into furnace gas
Long-life thermal conductivity cell that provides accurate & stable readings of H2
Separate digital meters with backlight
Rechargeable battery operation
Built-in sample pump, filter & flow meter
Weatherproof (WP) cabinet with clear Lexan cover
| |
|
Recorder outputs of 0-1VDC or 4-20 mA
Sample pre-cooler for hot samples
Condensate removal for wet applications
Suitcase (K) style cabinet available
AC only power option
O2 and H2 alarms with LED
Detachable/portable data logger
Dual range O2 (0-25.0% & 0-2.0%)(Model 341WP)
Special PPM O2 sensor (100-9999 PPM)(340L)
| |
|
On air for H2 zero and % O2 span
On analyzed mixture of % H2 in N2 for H2 span and O2 zero
On analyzed mixture of PPM O2 in N2 for PPM O2 span | | |
|
|
|