Method of Detection |
|
Temperature compensated thermal conductivity (T/C) cell
|
|
Ranges Available |
|
0-100% SF6 in air or in N2
|
|
Resolution |
|
0.1% of He
|
|
Accuracy & Repeatability |
|
±2% of full scale
|
|
Drift |
|
±1% of full scale per day (after calibration)
|
|
Response Time (T-90) |
|
10-15 seconds to 90% step change - not including sample transport time
|
|
Ambient Temperature Range |
|
12 to 50 ºC 55 to 120 ºF
|
|
Linearity |
|
±2% of F.S. on SF6 in Air
|
|
Size & Weight |
|
9 1/2" L x 6 1/2" H x 7" D @ 8 lbs (24 x 16.5 x 18 cm @ 3.6 kg)
|
|
Power |
|
AC/DC Operation. 115 VAC, 60 Hz for recharging (220 VAC, 50 Hz available)
|
|
Output Options |
|
4-20 mA or 0-1 VDC
|
|
Alarms |
|
High or low SF6 alarms (optional)
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|