								产品名称:
										430DPL-N4氢气和露点连续过程分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Temp-controlled T/C cell for H2; capacitive ceramic sensor for dew point
  |  
  |  
Ranges Available | 
  | 
0-2, 0-5, 0-10, 0-20, 0-50, and 0-100% H2 -148 to 68°F (-100 to 20°C) Dew Point
  |  
  |  
Resolution | 
  | 
±0.1% H2, ±1° dew point
  |  
  |  
Accuracy & Repeatability | 
  | 
±1% of full scale
  |  
  |  
Drift | 
  | 
Zero and Span ±1% of full scale per week
  |  
  |  
Response Time (T-90) | 
  | 
10-15 seconds to 90% step change
  |  
  |  
Ambient Temperature Range | 
  | 
32 to 122°F (0 to 49°C). lower temperatures (-5°F, -20°C) with cold weather package
  |  
  |  
Linearity | 
  | 
1% of full scale
  |  
  |  
Size & Weight | 
  | 
Dimensions will vary depending on enclosure style and options required
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz 220 VAC, 50 Hz
  |  
  |  
Output Options | 
  | 
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
  |  
  |  
Alarms | 
  | 
High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
  |  
  |  
Analyzer Type | 
  | 
H2 Analyzer
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
For continuous analysis of up to 100% hydrogen (H2) and low range dew point in process gas streams such as heat treating, hydrogen/nitrogen atmospheres and H2 cooled generators. 
 
 
  |    |  
 
 |  
- Best value continuous H2 and dew point analyzer
 
- Bright digital readout
 
- Fast response (T90-10-15 seconds)
 
- Easy to operate
 
- Long life thermal conductivity (T/C) H2 detector with high accuracy and repeatability.
 
- Cannot be burned out due to loss of flow
 
- Long life ceramic dew point sensor
 
- Low range of -148 to 68°F (-100 to 20°C) dew point
 
- Built in sample pump or pressure regulator
 
- Modular layout that is easy to maintain
  
 
 
  |    |  
 
 |  
- Hi/Low gas, low flow, and diagnostic alarms available
 
- Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
 
- Sample conditioning systems available for removal of acidic gas and dust
 
- Cabinet purge system available for use in hazardous areas
 
- One-step 'cal-now' button with touch screen LCD display
 
- Full automatic calibration with touch screen LCD display
 
- Cold weather package for operation to -5°F (-20°C)
 
- Cabinet coolers can be fitted to most models
 
- High temperature filters and probes
  
 
 
  |    |  
 
 |  
- Nitrogen for zero
 
- Analyzed calibration gas for H2 and known ambient dew point for dew point span.
   |    |    | 
							 
						  | 
					 | 
					
						 |