										206BT便携式低量程露点分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Long-life capacitance sensor
  |  
  |  
Ranges Available | 
  | 
-112°F to +68°F (-80°C to +20°C) Dew Point
  |  
  |  
Resolution | 
  | 
1°F (1°C) Dew Point
  |  
  |  
Accuracy & Repeatability | 
  | 
±1°C from -60 to +20°C Dew Point; ±2°C from -80 to -60°C ±1°F from -76 to +68°F Dew Point; ±2°F from -112 to -76°F
  |  
  |  
Drift | 
  | 
±2% of full scale per month maximum
  |  
  |  
Response Time (T-90) | 
  | 
Less than 30 seconds at higher dew points; lower dew points may take up to 1 minute (sensor must be in pre-dried state)
  |  
  |  
Ambient Temperature Range | 
  | 
-18 to 50 ºC 0 to 120 ºF
  |  
  |  
 
Linearity | 
  | 
±2°F/C over measurement range
  |  
  |  
Size & Weight | 
  | 
BT Style - Approx. 9" H x 8" W x 10" D @ 8 lbs (23 x 20 x 25 cm @ 3.6 kg) K Style - Approx. 14" L x 6" H x 10 1/2" D @ 12 lbs (35 x 15 x 26 cm @ 5.5 kg)
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz for recharging 220 VAC 50 Hz available
  |  
  |  
Output Options | 
  | 
0-1 V or 4-20 mA
  |  
  |  
Alarms | 
  | 
High or low dew point alarms with LED (optional)
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
For checking lower range dew points of air, hydrogen/nitrogen atmosphere furnace gases or other dry, oil-free process gases.
 
 
   |    |  
 
 |  
- Low range of -112 to +68°F (or -80 to 20°C) Dew Point
 
- Easy to operate
 
- Fast warm-up and response
 
- Use on any clean, non-corrosive industrial gases or air
 
- Long-life dew point sensor
 
- May be calibrated in the field
 
- Large LCD digital readout with backlight
 
- Configurations available for wide range of sample pressures
 
- No external adjustments to make
 
 
    |    |  
 
 |  
- Recorder outputs of 0-1 V or 4-20 mA
 
- Rugged "K" style weathertight suitcase enclosure
 
- high or low dew point alarm
  
 
 
  |    |  
 
 |  
- On analyzed dry nitrogen (N2) for low setting
 
- On known, current ambient air dew point for span
 
- Alternatively, the analyzer may be returned to Nova for a certified calibration
  
 
 
  |    |  
 
 |  
The 206 Series should not be used to measure dew points when trace amounts of sulfur trioxide (SO3), chlorine (Cl2), hydrogen chloride (HCl), or ammonia (NH3) are present.  |    |    | 
							 
						  | 
					 | 
					
						 |