								产品名称:
										256N4低量程连续露点分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Dielectric polymer sensor that changes capacitance with moisture variation
  |  
  |  
Ranges Available | 
  | 
-112°F to +68°F (-80°C to +20°C) Dew Point
  |  
  |  
Resolution | 
  | 
1°F (1°C) Dew Point
  |  
  |  
Accuracy & Repeatability | 
  | 
±1°C from -60° to +20°C Dew Point; ±2°C from -80° to -60°C Dew Point ±1°F from -76° to +68°F Dew Point; ±2°F from -112° to -76°F Dew Point
  |  
  |  
Drift | 
  | 
±2% of full scale per month maximum
  |  
  |  
Response Time (T-90) | 
  | 
Less than 60 seconds to T90 at higher dew points; Lower dew points may take an hour or more to dry down from ambient air
  |  
  |  
Ambient Temperature Range | 
  | 
-18 to 49 ºC 0 to 120 ºF
  |  
  |  
 
Linearity | 
  | 
±2°F / °C over measurement range
  |  
  |  
Size & Weight | 
  | 
16" W x 14" H x 8" D @ 10 lbs (41 x 36 x 20 cm @ 4.5 kg)
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz for recharging 220 VAC, 50 Hz available
  |  
  |  
Output Options | 
  | 
4-20 mA or 0-1 VDC
  |  
  |  
Alarms | 
  | 
High or low dew point alarms with LED (optional)
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
For checking lower range dew points of air, hydrogen/nitrogen atmospheres or other dry-oil free gases.
 
 
   |    |  
 
 |  
- Low range analysis available, down to -112° to +68°F (-80°C to +20°C) Dew Point
 
- Rugged design that is easy to operate
 
- Use on any non-corrosive industrial gases or oil-free air
 
- Long-life dielectric polymer sensor
 
- May be calibrated in the field
 
- Bright digital readout
 
- Built in sample pump, pressure regulator or filter
 
- 4-20 mA output, non-isolated
 
- NEMA 12 enclosure standard
  
 
 
  |    |  
 
 |  
- High or low dew point alarm with LED on panel
 
- 0-1VDC or isolated 4-20 mA recorder outputs
 
- Corrosion resistant wall mount enclosure
 
- NEMA 4, NEMA 4X and rack mount enclosures available
  
 
 
  |    |  
 
 |  
- On analyzed dry nitrogen for zero
 
- On known, current ambient air dew point for span
 
- Alternatively, the analyzer may be returned to Nova for a certified calibration
   |    |    | 
							 
						  | 
					 | 
					
						 |