								产品名称:
										320WP便携式烟气氧分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Long life electrochemical oxygen sensor
  |  
  |  
Ranges Available | 
  | 
0-5.0%, 0-10.0%, 0-25.0%, 0-50.0%
  |  
  |  
Resolution | 
  | 
0.1% O2
  |  
  |  
Accuracy & Repeatability | 
  | 
±1% of full scale (±2% of full scale above 30% O2)
  |  
  |  
Drift | 
  | 
< 2% of full scale for 8 hours of operation
  |  
  |  
Response Time (T-90) | 
  | 
8-10 seconds for 90% of step change
  |  
  |  
Ambient Temperature Range | 
  | 
0 to 40 ºC 32 to 105 ºF
  |  
  |  
 
Linearity | 
  | 
±1% of full scale
  |  
  |  
Size & Weight | 
  | 
10" L x 7 1/2" H x 6 1/2" D @ 8 lbs (25.5 x 19 x 16.5 cm @ 3.6 kg)
  |  
  |  
Power | 
  | 
AC/DC Operation, 115 VAC, 60 Hz for recharging (Other voltages available)
  |  
  |  
Output Options | 
  | 
0-1 V or 4-20 mA (optional)
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
Analysis of oxygen (O2) in flue gases from any type of fuel. For checking combustion efficiency, air infiltration and burner & control performance of furnaces, heaters and boilers. May be used in commercial, industrial and residential settings. 
 
 
  |    |  
 
 |  
- Rugged design that is easy to operate and maintain
 
- Fast warm-up and response
 
- Long life 'fuel cell' oxygen sensor (3-4 year life)
 
- Digital readout meter with backlight
 
- Rechargeable battery operation
 
- Built in sample pump, filter and flow meter
 
- Continuous condensate removal
 
- For use on any fuel
 
- Weatherproof (WP) cabinet with clear Lexan cover
 
- Stainless steel probe with sample hose
 
- Pays for itself in months through fuel savings
  
 
 
  |    |  
 
 |  
- Outputs of 0-1 V or 4-20 mA
 
- Special sampling probes
 
- Sample pre-cooler
 
- Suitcase (K) style cabinet available
 
- AC power only operation
 
- Detachable/portable data logger
  
 
 
  |    |  
 
 |  
- On air for span
 
- On nitrogen for zero
   |    |    | 
							 
						  | 
					 | 
					
						 |