|         
         | 
					
							
								|    | 
							 
							
								产品名称:
										510D缺氧报警监视器
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Customer replaceable electrochemical 'fuel cell' oxygen sensor, expected life 5 years
  |  
  |  
Ranges Available | 
  | 
0-30% O2
  |  
  |  
Resolution | 
  | 
0.1% O2
  |  
  |  
Accuracy & Repeatability | 
  | 
±2% of full scale
  |  
  |  
Drift | 
  | 
Less than 1% of full scale per month
  |  
  |  
Response Time (T-90) | 
  | 
6-8 seconds to 90% step change
  |  
  |  
Ambient Temperature Range | 
  | 
32 to 122°F (0 to 50°C), may be operated at lower temperature with auxiliary heater
  |  
  |  
Linearity | 
  | 
±2% of full scale
  |  
  |  
Size & Weight | 
  | 
10" H x 8" W x 6" D (25.4 x 20.3 x 15.2 cm) @ 6 lbs. (2.7 kg)
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz 12 VDC, 60 Hz 220 VAC, 50 Hz operation 24 VDC, 60 Hz
  |  
  |  
Output Options | 
  | 
0-1 VDC 4-20 mA
  |  
  |  
Alarms | 
  | 
High and low SPDT contacts with 10 A, 115 VAC rating; Two alarm levels standard; 'Sonalert' audible alarm with reset button; Indicator light for each alarm level; Strobe light and 'fail safe' alarms available
  |  
  |  
Analyzer Type | 
  | 
O2 Gas Analyzer
  |  
  |   
  |    |  
 
 |  
To monitor confined spaces, labs, control rooms, tunnels, oxygen plants, etc. for oxygen (O2) deficiency or excess. 
 
 
  |    |  
 
 |  
- Best value high accuracy O2 monitor
 
- Bright digital readout
 
- 5 year life electrochemical O2 sensor
 
- Two levels of O2 alarm with relay contacts and 'Sonalert' audible alarm with 'Acknowledge' button
 
- Easy to use and calibrate
 
- Fast response (T90 - 6 to 8 seconds)
 
- Corrosion resistant NEMA 12 enclosure
 
- 4-20 mA or 0-1 VDC recorder output
 
- Modular layout that is easy to maintain
  
 
 
  |    |  
 
 |  
- 'Fail Safe' alarms available
 
- Low flow alarm on pumped sample models
 
- NEMA 4X enclosure
 
- Strobe light
  
 
 
  |    |  
 
 |  
- Although calibration of zero should not be necessary, it can be checked by flowing 1 LPM of nitrogen through the analyzer
 
- Analyzed gas or ambient air with 20.9% O2 for span
   |    |    | 
							 
						  | 
					 | 
					
						 |