								产品名称:
										413N4氧气过程分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Customer replaceable zirconia O2 sensor. Expected life 5 years or more.
  |  
  |  
Ranges Available | 
  | 
413 - any range from 0-2.0% to 0-96% O2 413L - any range from 0-200 PPM to 0-20000 PPM O2
  |  
  |  
Resolution | 
  | 
0.1% O2 or % versions; 1 PPM O2 on PPM versions
  |  
  |  
Accuracy & Repeatability | 
  | 
±0.5% of full scale
  |  
  |  
Drift | 
  | 
Less than 0.3% of full scale per month
  |  
  |  
Response Time (T-90) | 
  | 
2-3 seconds to 90% step change
  |  
  |  
Ambient Temperature Range | 
  | 
32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package
  |  
  |  
Linearity | 
  | 
±0.5% of full scale
  |  
  |  
Size & Weight | 
  | 
Dimensions will vary depending on enclosure style and options required
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz 220 VAC, 50 Hz
  |  
  |  
Output Options | 
  | 
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
  |  
  |  
Alarms | 
  | 
High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
  |  
  |  
Analyzer Type | 
  | 
O2 Analyzer
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
For continuous analysis of PPM to 100% oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, oxygen deficiency analysis, and many more.
 
 
   |    |  
 
 |  
- High accuracy continuous O2 analyzer 
 - Bright digital readout, and 4 - 20 mA outputs 
 - Long life zirconia O2 sensor 
 - Fast response (T90 - 2 to 3 seconds) 
 - Microprocessor control for push button calibration 
 - Easy-to-maintain modular layout 
 - Built in sample pump or pressure regulator
  
 
 
  |    |  
 
 |  
- Hi/Low gas, low flow, and diagnostic alarms available 
 - Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available 
 - Sample conditioning systems available for removal of moisture, acidic gas, and dust 
 - Cabinet purge system available for use in hazardous areas 
 - One-step 'cal-now' button with touch screen LCD display 
 - Full automatic calibration with touch screen LCD display 
 - Cold weather package for operation to -5°F (-20°C) 
 - Cabinet coolers can be fitted to most models 
 - High temperature filters and probes
  
 
 
  |    |  
 
 |  
- Nitrogen for zero 
 - Air or analyzed calibration gas for span
   |    |    | 
							 
						  | 
					 | 
					
						 |