								产品名称:
										412RM氧气过程分析仪
										   | 
							 
							
								 
									  
									
| 
 |  
Method of Detection | 
  | 
Magneto-dynamic paramagnetic O2 cell
  |  
  |  
Ranges Available | 
  | 
412 - any range from 0-2.0% to 0-100.0% O2; 412HP - 98.0 to 100.0 or 0 to 100.00 O2
  |  
  |  
Resolution | 
  | 
412 - 0.1% O2; 412HP - 0.01% O2
  |  
  |  
Accuracy & Repeatability | 
  | 
±0.5% of full scale or 0.05%, whichever is greater
  |  
  |  
Drift | 
  | 
±0.2% of full scale per month
  |  
  |  
Response Time (T-90) | 
  | 
4-5 seconds
  |  
  |  
Ambient Temperature Range | 
  | 
32 to 122°F (0 to 50°C) @ 5-95% RH non-condensing. Lower temperatures (-5°F, -20°C) with Cold Weather Package. Sample dew point should be at least 9°F (5°C) below lowest sample temperature.
  |  
  |  
Linearity | 
  | 
±0.5% of full scale or 0.05%, whichever is greater
  |  
  |  
Size & Weight | 
  | 
Dimensions will vary depending on enclosure style and options required
  |  
  |  
Power | 
  | 
115 VAC, 60 Hz 220 VAC, 50 Hz
  |  
  |  
Output Options | 
  | 
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
  |  
  |  
Alarms | 
  | 
High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
  |  
  |  
Analyzer Type | 
  | 
O2 Analyzer
  |  
  |  
Hazardous Area Rating Options | 
  | 
General Purpose, Non-Hazardous
  |  
  |   
  |    |  
 
 |  
For continuous analysis of oxygen (O2) in process gas streams that contain gases that interfere with other methods of O2 measurement. For high accuracy or high purity applications.
 
 
   |    |  
 
 |  
- Highest accuracy O2 analyzer, very stable 
 - Fast response (T90 - 4 to 5 seconds) and bright digital readout 
 - Linear over entire 0-100% range 
 - Microprocessor control for push button calibration 
 - Non-consumable O2 detector, no periodic replacement required 
 - Built in sample pump or pressure regulator, and 4 - 20 mA outputs 
 - To improve accuracy, detector is mounted in temperature controlled heated box and compensated for changes in barometric pressure at the vent
  
 
 
  |    |  
 
 |  
- Hi/Low gas, low flow, and diagnostic alarms available 
 - Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available 
 - Sample conditioning systems available for removal of moisture, acidic gas, and dust 
 - Cabinet purge system available for use in hazardous areas 
 - One-step 'cal-now' button with touch screen LCD display 
 - Full automatic calibration with touch screen LCD display 
 - Cold weather package for operation to -20°C (-4°F) 
 - Cabinet coolers can be fitted to most models 
 - High purity version available for 98-100% O2 (412HP)
  
 
 
  |    |  
 
 |  
- Nitrogen for zero 
 - Air or analyzed calibration gas for span
   |    |    | 
							 
						  | 
					 | 
					
						 |