Method of Detection |
|
Customer replaceable electrochemical oxygen and CO sensors Solid state infra-red detector for CO2
|
|
Ranges Available |
|
0-1800°F (0-1000°C) stack temperature 0-20.0% CO2 0-2000 PPM and 0-4.00% CO - Switch selectable 0-25.0% Oxygen
|
|
Resolution |
|
0.1% O2, 1 PPM on CO and CO2
|
|
Accuracy & Repeatability |
|
2% of full scale for O2 and CO2; ±10 PPM CO
|
|
Drift |
|
<2% full scale per 8 hours of continuous operation
|
|
Response Time (T-90) |
|
10-15 seconds for O2; 20-30 seconds for CO and CO2
|
|
Ambient Temperature Range |
|
0 to 40 ºC 32 to 105 ºF
|
|
Linearity |
|
±1.0% of full scale for each gas measured
|
|
Size & Weight |
|
18" L x 7" H x 12" D @ 15 lbs (45.7 x 30.5 x 17.8 cm @ 6.8 kg)
|
|
Power |
|
AC/DC Operation, 115 VAC 60 Hz for recharging (Other voltages available.)
|
|
Output Options |
|
0-1 VDC 4-20 mA
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
Analysis of oxygen (O2), carbon monoxide (CO), and carbon dioxide (CO2). For checking the combustion efficiency and burner & control performance of furnaces, heaters, and boilers. May be used in commercial, industrial, and residential settings.
| |
|
- Rugged design that is easy to operate
- Fast warm up and response
- CO2 actually measured by infrared detector, not calculated
- Dual CO range : 0-2000 ppm & 0-4.00%
- Rapid reading recovery after CO 'overdose'
- Digital readout meters with backlight
- Modular design that is easy to maintain
- Rechargeable battery operation
- Built in sample pump, filter and flow meter
- Active condensate removal
- Weatherproof (WP) cabinet with clear Lexan cover
- Stainless steel probe with sample hose
| |
|
- Outputs of 0-1 V or 4-20 mA (can be isolated)
- Stack temperature readout (375T)
- Special sampling probes
- Suitcase (K) style cabinet available
- AC power only operation
- Gas alarms with LED
- Detachable/portable data logger
| |
|
- Air for O2 span and CO/CO2 zero
- Analyzed gas mixtures of CO2 and CO in nitrogen for span
| |
|
|
|
|