Method of Detection |
|
Temperature compensated thermal conductivity (T/C) cell
|
|
Ranges Available |
|
0-100% H2 in a binary gas mixture
|
|
Resolution |
|
0.1% of H2
|
|
Accuracy & Repeatability |
|
±2% of full scale
|
|
Drift |
|
±1% of full scale per week
|
|
Response Time (T-90) |
|
10-15 seconds to 90% step change - not including sample transport time
|
|
Ambient Temperature Range |
|
12 to 50 ºC 55 to 120 ºF
|
|
Linearity |
|
±2% of F.S.
|
|
Size & Weight |
|
10" L x 7 1/2" H x 6 1/2" D @ 8 lbs (25.5 x 19 x 16.5 cm @ 3.6 kg)
|
|
Power |
|
AC/DC Operation. 115 VAC, 60 Hz for recharging (220 VAC, 50 Hz available)
|
|
Output Options |
|
0-1 VDC 4-20 mA
|
|
Alarms |
|
H2 alarm with LED (optional)
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
For analysis of hydrogen (H2) in a binary gas mixture in process gases such as H2 in air, H2 in nitrogen (N2), H2 in carbon dioxide (CO2), H2 in argon (Ar), H2 in oxygen (O2), etc.
| |
|
- Rugged design that is easy to operate
- Fast warm up and response
- Long life thermal conductivity cell that provides accurate and stable readings
- Digital meter readout with backlight
- Modular layout that is easy to maintain
- Rechargeable 'gel cell' battery operated
- Built in sample pump or flow regulator
- Weatherproof (WP) cabinet with clear lexan cover
| |
|
- Recorder outputs of 0-1 V or 4-20 mA
- Sample pre-cooler for hot samples
- Condensate removal for wet applications
- Suitcase (K) style cabinet available
- AC power only operation
- H2 alarm with LED
- Detachable/portable data logger
| |
|
- Ambient air for zero
- Gas cylinder of known H2 for span
| |
|
|
|
|