Method of Detection |
|
Microprocessor based infrared detector for CO2
|
|
Ranges Available |
|
0-20% CO2
|
|
Resolution |
|
0.1% CO2
|
|
Accuracy & Repeatability |
|
Within ±2% full scale
|
|
Drift |
|
Within 1% full scale per 8 hours of continuous operation
|
|
Response Time (T-90) |
|
30 - 40 seconds
|
|
Ambient Temperature Range |
|
0 to 40 ºC 32 to 105 ºF
|
|
Linearity |
|
±1.0% of full scale
|
|
Size & Weight |
|
14" L x 10 1/2" W x 6" H @ 8 lbs (36 x 27 x 15 cm @ 5.5 kg)
|
|
Power |
|
AC/DC Operation, 115 VAC, 60 Hz for recharging
|
|
Output Options |
|
0-1 VDC 4-20 mA
|
|
Alarms |
|
High or low CO2 alarm with LED (optional)
|
|
Hazardous Area Rating Options |
|
General Purpose, Non-Hazardous
|
|
| |
|
Analysis of carbon dioxide (CO2). For checking combustion efficiency, and burner & control performance of furnaces, heaters and boilers. May be used in commercial, industrial and residential settings.
| |
|
- Rugged design that is easy to operate
- Fast warm-up and response
- Solid state infrared detector
- Digital readout meter with backlight
- Modular layout that is easy to maintain
- Rechargeable battery operation
- Built in sample pump, filter and flowmeter
- Active condensate removal
- Weatherproof (WP) cabinet with clear Lexan cover
- Stainless steel probe with sample hose
| |
|
- Recorder outputs of 0-1 V or 4-20 mA (can be isolated)
- Stack temperature readout (302T)
- Special sampling probes
- Sample pre-cooler
- Suitcase (K) style cabinet available
- AC power only operation
- CO2 alarms with LED
- Data logger
| |
|
- Air for zero (except low range models which have a CO2 scrubber for zeroing).
- Analyzed CO2 cal gas for span.
| |