Specifications

Method of Detection

Single cell NDIR with no moving parts

Ranges Available

481 - 0-1, 0-5, 0-10, 0-20, 0-30, 0-50, 0-100% CO

Resolution

0.1% CO on % versions, 1 PPM CO on PPM versions

Accuracy & Repeatability

1% of full scale

Drift

Less than 2% of full scale per month

Response Time (T-90)

Less than 15 seconds

Ambient Temperature Range

40 to 122°F (5 to 50°C), Lower temperatures (-5°C, -20°F) with Cold Weather Package

Linearity

±1% of full scale

Size & Weight

Physical data will vary depending on enclosure style and options required

Power

115 VAC, 60 Hz
220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating
Low flow alarm

Analyzer Type

CO Analyzer

Hazardous Area Rating Options

General Purpose, Non-Hazardous


Applications
    For continuous analysis of carbon Monoxide (CO) in process gas streams using an infrared detector. Percent and PPM levels available as Models 481 and 481L respectively.




Features
  • Percent or PPM levels CO by infrared
  • Detector is sensitive, very reliable, and non-depleting
  • Bright digital readout, 4 - 20 mA outputs
  • Fast response (T90 - less than 15 seconds)
  • Microprocessor control for push button calibration
  • Easy-to-maintain modular layout
  • Built in sample pump, flowmeter and filters




Options
  • Hi/Low gas, low flow, and diagnostic alarms available
  • Isolated analog, RS232, RS485, MODBUS and Ethernet outputs available
  • Sample conditioning systems available for removal of moisture, acidic gas and dust
  • Cabinet purge system available for use in hazardous areas
  • One-step 'cal-now' button with touch screen LCD display
  • Full automatic calibration with touch screen LCD display
  • Cold weather package for operation to -20°C (-5 °F)
  • Cabinet coolers can be fitted to most models
  • Sensors mounted in temperature controlled heated box to improve accuracy.
  • High temperature filters and probes




Calibration
  • Easily zero checked on ambient air
  • Span using a known CO in nitrogen (N2) calibration gas.