|
|
Method of Detection |
|
Customer replaceable 'fuel cell' oxygen sensor
|
|
Ranges Available |
|
Any range between 0-10 PPM and 0-10000 PPM
|
|
Resolution |
|
±1 PPM
|
|
Accuracy & Repeatability |
|
±1% of full scale (minimum 0.5 PPM)
|
|
Drift |
|
Less than 1% of full scale per week
|
|
Response Time (T-90) |
|
O2 sensor may require an initial purge down from air to low PPM O2 which may last up to 6 hours. After purge, T90 response from 1000 to 0 PPM O2 is 30 seconds.
|
|
Ambient Temperature Range |
|
32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package
|
|
Linearity |
|
±1% of full scale (minimum 0.5 PPM)
|
|
Size & Weight |
|
Dimensions will vary depending on enclosure style and options required
|
|
Power |
|
115 VAC, 60 Hz 220 VAC, 50 Hz
|
|
Output Options |
|
4-20 mA into 500 ohms non-isolated standard Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
|
|
Alarms |
|
High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating Low flow alarm
|
|
Analyzer Type |
|
O2 Analyzer
|
|
Hazardous Area Rating Options |
|
Class 1, Div 1, Grp A-D Hazardous Area Rating
|
|
| |
|
For continuous analysis of trace PPM levels of oxygen (O2) in process gas streams. Applications include blanketing mixtures, glove boxes, electronics, research, hydrogen/nitrogen mixtures, and many others.
| |
|
- Bright digital readout in PPM O2
- Ranges available from 0 to 10 PPM to 0 to 10000 PPM O2
- Microprocessor control for push button calibration
- Unaffected by hydrocarbons in the sample
- Easy-to-maintain, modular layout
- Built in sample pump or pressure regulator, and 4-20 mA outputs
- Manual or solenoid valves trap nitrogen (N2) in cell to avoid depletion when analyzer is powered off
| |
|
- Hi/Low gas, low flow, and diagnostic alarms available
- Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
- Sample conditioning systems available for removal of moisture, acidic gas, and dust
- Cabinet purge system available for use in hazardous areas
- One-step 'cal-now' button with touch screen LCD display
- Full automatic calibration with touch screen LCD display
- Cold weather package for operation to -20°C (-4°F)
- Cabinet coolers can be fitted to most models
- Sensors mounted in temperature controlled heated box to improve accuracy
| |
|
- Nitrogen for zero
- Air or analyzed calibration gas for span
| |
|
|
|
|